Liquistation CSF39
Technical data
Hauser
183
> 0.6 m/s (> 1.9 ft/s) for 10 mm (3/8") ID, in accordance with Ö 5893; US EPA
Suction height
•
Vacuum pump:
Max. 6 m (20 ft) or max. 8 m (26 ft), depending on the version
•
Peristaltic pump:
Max. 8 m (26 ft)
Hose length
Max. 30 m (98 ft)
Temperature control
Temperature sensors:
• Sampling compartment temperature
• Sample temperature (optional)
• Outside temperature (optional)
Cooling module:
• Sample temperature range: 2 to 20 °C (36 to 68 °F)
Factory setting: 4 °C (39 °F)
• Automatic defrost system
• Cooling rate in accordance with Ö 5893 (Austrian standard):
4 liters of water at 20 °C cool down to 4 °C in less than 210 minutes
• Temperature constancy of sample at 4 °C over the operating temperature range of -15 to
40 °C (5 to 105 °F)
15.11 Environment
Ambient temperature
range
With cooling module:
-20 to 40 °C (0 to 104 °F)
Without cooling module:
0 to 40 °C (32 to 104 °F)
With ASA+PC or stainless steel
housing:
-20 to 40 °C (0 to 104 °F)
With plastic polystyrene housing:
0 to 40 °C (32 to 104 °F)
Storage temperature
–20 to 60 °C (–4 to 140 °F)
Electrical safety
In accordance with EN 61010-1, protection class I, environment ≤ 2000 m (6500 ft) above
MSL. The device is designed for pollution degree 2.
Humidity
10 to 95%, not condensing
Degree of protection
• Front dosing compartment: IP 54
• Rear dosing compartment: IP 33
• Front panel with display (internal): IP 65
• Sample compartment: IP 54
Electromagnetic
compatibility
Interference emission and interference immunity as per EN 61326-1:2013, Class A for
Industry