Liquistation CSF39
Diagnostics and troubleshooting
Hauser
145
Entry
Example
Info
Running sample number
1
Running sample number in the current
program
Flow sum since last sampling
1
For
flow-paced
and
time/flow-paced
sampling:
--> Flow since the last sampling
For all other types of sampling:
--> Display: 0
11.9.4 Bottle statistics
The bottle statistics of the sampler are displayed:
‣
In
Menu
/
Diagnostics
/
Logbooks
/
Logbook program
select the menu item
Show
summary of current program
, or simply select the STAT soft key while the sampling
program is running
The statistics are displayed for each individual bottle when the program is started.
This gives you detailed feedback on the last sampling operations.
The statistics are deleted when the following event occurs:
Program is started
The statistics are selectively overwritten when the following event occurs:
When the first bottle is reached in situations where "Continuous operation" is
configured as the end of the program in the program settings
The statistics are displayed as follows:
A0036867-EN
User interface
Info
bt
The bottle number is displayed.
hh:mm
The time the first sample was transferred to the bottle is displayed.
DD-hh:mm
The time the first sample was transferred to the bottle is displayed.
Smp
Displays how often sampling was triggered per bottle.
n.s
Indicates the number of times a sample was not taken even though sampling was
triggered. This can occur if the maximum permissible fill volume for the bottle has
been reached but the system is still supposed to transfer samples to the bottle. The
"Overfill sensor" message is displayed while the program is active.
n.f
The value indicates how often sampling was canceled since the system was unable to
take in any medium, or enough medium, into the dosing chamber to cover the LF1
probe.
ml
The sampling volume collected per bottle is displayed.
Q
The total flow for every bottle is displayed (if connected).
Summary of Contents for Liquistation CSF39
Page 2: ......
Page 188: ...www addresses endress com ...