Diagnostics and troubleshooting
Liquistation CSF39
140
Hauser
11.6.2 Sensor-specific diagnostics messages
Operating Instructions "Memosens", BA01245C
11.7 Pending diagnostic messages
The Diagnostics menu contains all the information on the device status.
Furthermore, various service functions are available.
The following messages are directly displayed every time you enter the menu:
•
Most important message
Diagnostic message recorded with the highest criticality level
•
Past message
Diagnostic message whose cause is no longer present.
All the other functions in the Diagnostics menu are described in the following chapters.
Diagnostic messages associated with sampling are deleted under the following conditions:
• Diagnostic messages caused by sampling are deleted automatically with the next
successful sampling.
• Diagnostic messages caused by the level of medium in the bottle are deleted the next
time the bottle is changed.
If the diagnostic message "M313 liquid sensor" appears 5 times in succession when
executing a program, the active program is aborted for reasons of safety. This
behavior on the part of the device cannot be altered by deactivating the diagnostic
message under
Menu
/
Setup
/
General settings
/
Extended setup
/
Diagnostics
settings
.
11.8 Diagnosis list
All the current diagnostic messages are listed here.
A time stamp is available for each message. Furthermore, the user also sees the
configuration and description of the message as saved in
Menu
/
Setup
/
General settings
/
Extended setup
/
Diagnostics settings
/
Diag. behavior
.
11.9 Logbooks
11.9.1 Available logbooks
Types of logbooks
• Logbooks physically available (all apart from the overall logbook)
• Database view of all logbooks (=overall logbook)
Logbook
Visible in
Max.
entries
Can be
disabled
Logbook
can be
deleted
Entries
can be
deleted
Can be
exported
Overall logbook
All events
20000
Yes
No
Yes
No
Calibration logbook
Calibration events
75
(Yes)
No
Yes
Yes
Operation logbook
Configuration events
250
(Yes)
No
Yes
Yes
Diagnostics logbook
Diagnostic events
250
(Yes)
No
Yes
Yes
Program log
Program logbook
5000
Yes
No
Yes
Yes
Version logbook
All events
50
No
No
No
Yes
Hardware version
logbook
All events
125
No
No
No
Yes