Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
NOTICE
The device does not have a power switch
‣
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the
local regulations for installation.
‣
Use a HBC fuse with 10 A, 250 V AC for samplers with a CSA approval
‣
The circuit breaker must be a switch or power switch, and you must label it as the
circuit breaker for the device.
‣
The protective ground connection must be established before all other connections. If
the protective ground is disconnected, this can be a source of danger.
‣
A circuit breaker must be located near the device.
‣
For 24V versions, the power supply at the voltage source must be isolated from cables
carrying low voltage (110/230V AC) by double or reinforced insulation.
Operation with non-stationary mains cable connection to sampler (optional)
6.1
Connecting the sensors
6.1.1
Connection compartment in the controller housing
A0012843
1
2
A0042244
1
1 E base module
2
Sampler controller
The controller housing has a separate connection
compartment. Release the six cover screws to open the
connection compartment:
‣
Release 6 cover screws with a Phillips screwdriver
to open the display cover.
Display cover open, version with base module E