Liquiline System CA80TN
Commissioning
Hauser
41
9.2
Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions!
‣
Check that all connections have been established correctly in accordance with the
wiring diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.
L
WARNING
Connection errors
The safety of people and of the measuring point is under threat. The manufacturer does
not accept any responsibility for errors that result from failure to comply with the
instructions in this manual.
‣
Put the device into operation only if you can answer
yes
to
all
the following questions.
Device condition and specifications
‣
Are the hoses free from damage on the outside?
Pressure reactor
‣
Have all of the reactor's connections been installed correctly?
‣
Has the reactor's safety cover been installed?
Visual inspection of the liquid-bearing lines
‣
Check the hose connections using the hose connection diagram.
‣
Are all the hose connections leak-tight?
‣
is the sample hose in the hose gland strain-relieved?
‣
Have the bottles with reagents, and standard been inserted and connected?
9.3
Switching on the measuring device
L
WARNING
The strobe lamp produces high-intensity visible and invisible light radiation
Can cause severe damage to eyes and skin!
‣
Never look into the strobe lamp directly.
‣
Do not switch on the device if the reactor or the sensor unit are not installed or are
damaged.
‣
For maintenance work, make sure to always de-energize the device.
1. Connect the power supply.
2. Wait for the initialization to finish.
9.4
Access to the configuration (only versions CA80TN-
HR)
Enabling access to the configuration
Only authorized persons may make settings on the device. Access is password-protected.
1. Call up:
MENU
/
Enter maintenance level
.
2. Enter 8888 as the password.
Access is enabled and you can change the settings.
3. Change the password to a new, secure password:
MENU
/
General settings
/
Extended
setup
/
Data management
/
Change maintenance level password
.
If you forget your password, you can reset it here using the PUK supplied:
Reset password
with PUK
.