Liquiline System CA80TN
Operation
Hauser
109
2. Configure the
Formula
mathematical function:
Source A
= input signal phosphate
and
Source B
= input signal volume flow.
Formula:
A*B*x
(where x is an application-specific proportionality factor)
3. Select this formula as the source e.g. of the current output or of a modulated binary
output.
4. Connect the valve or pump.
10.7.6 Diagnostic modules
Here you can configure a maximum of 8 individual diagnostic messages.
A diagnostic module has the following properties:
• The feed source can be configured like a binary output (relay, digital output).
• You can choose whether the diagnostic message should be output at a high level or low
level.
• You decide which error category (Namur class) should be assigned to the message.
• You can define a customized text to be output as the diagnostic message text.
In addition you can switch off the factory diagnostic code for limit switches. This allows
you to:
• Use the limit switch on a purely functional basis (without a message)
• Configure application-specific message texts
• Control diagnostic modules directly by a digital signal or via a limit switch output
(enables the use of switch-on/switch-off delay, for example).
Menu/Setup/Additional functions/Diagnostic modules
Function
Options
Info
Diagnostic module 1 (961) ...
8 (968)
Source of data
Selection
• None
• Fieldbus signals
• Binary inputs
• Limit switches
Factory setting
None
Specify the input which should be the source of
data for the diagnostic message.
Measured value
Selection
depend on
Source of data
Factory setting
None
Specify the measured value that should trigger
the diagnostic message.
You can use different measured values
depending on the data source.
Active low
Selection
• Off
• On
Factory setting
On
On
: The output value is equal to the inverse
output value.
Short text
Free text
Give the diagnostic message a name.
Diagnostic module
assignment view
Shows an overview of the diagnostic modules
used.