Liquiline System CA80PH
Operation
Hauser
131
The pneumatic retractable assembly, e.g. CPA875, is activated by compressed air via a two-way
valve. As a result, the assembly assumes either the position 0 ("Measure" - sensor in medium) or
position 1 ("Service" - sensor in rinse chamber). Media such as water, acids or bases are supplied
via valves or pumps. Here there are two states 0 (= "off" or "closed") and 1 (= "on" or "open").
The hardware required for "Chemoclean Plus", such as control valves, pumps, compressed
air supply, media etc., must be provided by the customer.
Defining the cleaning cycle
Path: Menu/Setup/Additional functions/Cleaning/Cleaning 1 ... 4
Function
Options
Info
Cleaning cycle
Options
• Off
• Interval
• Weekly program
Factory setting
Off
Choose from a cleaning routine that is restarted at set
intervals and a user-definable weekly program.
Cleaning cycle="Interval"
Cleaning interval
00-00:01 to 07-00:00
(DD-hh:mm)
Factory setting
01-00:00
The interval value can be between 1 minute and 7 days.
Example: You have set the value "01-00:00". Each day,
the cleaning cycle starts at the same time you started the
first cleaning cycle.
Cleaning cycle="Weekly program"
Daily event times
00:00 to 23:59
(HH:MM)
1. "Event time 1 to 6": Define up to 6 times.
You can then choose from these for each
weekday afterwards.
2. For each day of the week, individually choose which
of the 6 times is to be used for a cleaning routine on
this particular day.
In this way you create weekly programs that are perfec-
tly adapted to your process.
Weekdays
Monday ... Sunday