Liquicap M FMI52 HART
Mounting
Hauser
17
gn/ye
ye
rd
bk
6
1
2
3
4
5
A0040734
7
Cable connections
1
External screening (not required)
2
Strand black (bk) (not required)
3
Coaxial cable with central core and screening
4
Solder the red (rd) strand with the central core of the coaxial cable (probe)
5
Solder the strand with the screening of the yellow (ye) coaxial cable (ground)
6
Strand yellow and green (gn/ye) with a ring terminal
• We recommend reusing all strands with ring terminals in case of shortening the
connecting cable
• To avoid the risk of short-circuiting when the strands are not to be reused, the
connections of the new ring terminals must be insulated with a heat shrinking sleeve
• Use heat-shrink tubes to insulate all soldered joints
4.3
Post-installation check
After installing the measuring device, carry out the following checks:
Do a visual check for damages.
Does the device meet the specifications at the measuring point with regard to process
temperature and pressure, ambient temperature, measuring range?
Has the process connection been tightened with the tightening torque?
Check if the measuring points are correctly labeled.
Is the device adequately protected against precipitation and direct sunlight?