Levelflex FMP56, FMP57 FOUNDATION Fieldbus
Operating menu
Hauser
197
Logging interval
Navigation
Diagnostics → Data logging → Logging interval
Description
Define logging interval t
log
.
User entry
1.0 to 3
600.0 s
Factory setting
30.0 s
Additional information
This parameter defines the interval between the individual data points in the data log, and
thus the maximum loggable process time T
log
:
• If 1 logging channel is used: T
log
= 1000 ⋅ t
log
• If 2 logging channels are used: T
log
= 500 ⋅ t
log
• If 3 logging channels are used: T
log
= 333 ⋅ t
log
• If 4 logging channels are used: T
log
= 250 ⋅ t
log
Once this time elapses, the oldest data points in the data log are cyclically overwritten such
that a time of T
log
always remains in the memory (ring memory principle).
The logged data are deleted if this parameter is changed.
Example
When using 1 logging channel
• T
log
= 1000 ⋅ 1 s = 1
000 s ≈ 16.5 min
• T
log
= 1000 ⋅ 10 s = 1
000 s ≈ 2.75 h
• T
log
= 1000 ⋅ 80 s = 80
000 s ≈ 22 h
• T
log
= 1000 ⋅ 3
600 s = 3
600
000 s ≈ 41 d
Clear logging data
Navigation
Diagnostics → Data logging → Clear logging
Description
Initiate a deletion of the complete logging memory.
Selection
• Cancel
• Clear data
Factory setting
Cancel