Operating menu
Levelflex FMP55 HART
154
Hauser
Interface linearized
Navigation
Setup → Advanced setup → Linearization → Interface linearized
Prerequisite
Operating mode (→ 128)
Interface
or
Interface with capacitance
Description
Displays the linearized interface height.
Additional information
The unit is defined in the
Unit after linearization
Maximum value
Navigation
Setup → Advanced setup → Linearization → Maximum value
Prerequisite
Linearization type (→ 151)
has one of the following values:
• Linear
• Pyramid bottom
• Conical bottom
• Angled bottom
• Horizontal cylinder
• Sphere
Description
Linearized value corresponding to a level of 100%.
User entry
–50
000.0 to 50
000.0 %
Factory setting
100.0 %
Diameter
Navigation
Setup → Advanced setup → Linearization → Diameter
Prerequisite
Linearization type (→ 151)
has one of the following values:
• Horizontal cylinder
• Sphere
Description
Diameter of the cylindrical or spherical tank.
User entry
0 to 9
999.999 m
Factory setting
2 m
Additional information
The unit is defined in the
Distance unit