Levelflex FMP55 HART
Operating menu
Hauser
191
Level signal
Navigation
Diagnostics → Device check → Level signal
Prerequisite
Device check has been performed.
Description
Displays result of the device check for the level signal.
User interface
• Check not done
• Check not OK
• Check OK
Additional information
For
Level signal
=
Check not OK
: Check the mounting position of the device and the
dielectric constant of the medium.
Launch signal
Navigation
Diagnostics → Device check → Launch signal
Prerequisite
Device check has been performed.
Description
Displays result of the display check for the launch signal.
User interface
• Check not done
• Check not OK
• Check OK
Additional information
For
Launch signal
=
Check not OK
: Check the mounting position of the device. In non-
metallic vessels use a metal plate or a metal flange.
Interface signal
Navigation
Diagnostics → Device check → Interface signal
Prerequisite
•
=
Interface
or
Interface with capacitance
• Device check has been performed.
Description
Displays result of the device check for the interface signal.
User interface
• Check not done
• Check not OK
• Check OK