iTHERM TrustSens TM371, TM372
Operating menu and parameter description
Hauser
79
20 mA value
Navigation
Setup → Upper range value
Expert → Output → 20 mA value
Description
Use this function to assign a measured value to the current value 20 mA.
Factory setting
150 °C
Failure mode
Navigation
Setup → Failure mode
Expert → Output → Failure mode
Description
Use this function to select the signal on alarm level of the current output in an event of an
error.
Options
• High alarm
• Low alarm
Factory setting
Low alarm
14.2 Calibration menu
All the information that describes the self-calibration procedure as well as the online
wizard creating a calibration report.
Number of self-calibrations
Navigation
Calibration → Number of self-calibrations
Description
This counter displays the amount of all executed self-calibrations. It cannot be reset.
Stored self-calibration points
Navigation
Calibration → Stored self-calibration points
Description
Displays the amount of all stored self-calibration points. This device is able to store 350
self-calibration points. As soon as the memory would reach its limit, the oldest self-
calibration point will be overwritten.
Display
0 to 350
Summary of Contents for iTHERM TrustSens TM371
Page 2: ......
Page 106: ...www addresses endress com ...