Technical data
Proline Promag P 500 HART
188
Hauser
0
0
0
0
-20
-40
-40
20
20
40
100
60
140
[°F] [°C]
40 60 80 100
100
120 140 160
200
300
180
360
-20
-40
-40
T
A
T
F
[°F]
[°C]
1
A0029808
51
PTFE
T
A
Ambient temperature range
T
F
Fluid temperature
1
Colored area: the ambient temperature range of –10 to –40 °C (+14 to –40 °F) applies to stainless flanges
only
Conductivity
≥ 5 μS/cm for liquids in general. Stronger filter damping is required for very low
conductivity values.
Proline 500
The necessary minimum conductivity also depends on the cable length .
Pressure-temperature
ratings
An overview of the pressure-temperature ratings for the process connections is
provided in the "Technical Information" document
Pressure tightness
"–" = no specifications possible
Liner: PFA
Nominal diameter
Limit values for absolute pressure in [mbar] ([psi]) for medium temperatures:
[mm]
[in]
+25 °C (+77 °F)
+80 °C (+176 °F)
+100 to +180 °C
(+212 to +356 °F)
25
1
0 (0)
0 (0)
0 (0)
32
–
0 (0)
0 (0)
0 (0)
40
1 ½
0 (0)
0 (0)
0 (0)
50
2
0 (0)
0 (0)
0 (0)
65
–
0 (0)
0 (0)
0 (0)
80
3
0 (0)
0 (0)
0 (0)
100
4
0 (0)
0 (0)
0 (0)
125
–
0 (0)
0 (0)
0 (0)
150
6
0 (0)
0 (0)
0 (0)
200
8
0 (0)
0 (0)
0 (0)
Summary of Contents for Hart Proline Promag P 500
Page 210: ...www addresses endress com ...