Deltabar FMD72
Installation
15
6
Installation
When measuring in media containing solids, such as dirty liquids, installing separators and drain
valves is useful for capturing and removing sediment.
6.1
Mounting dimensions
For dimensions, see the "Mechanical construction" section in TI01033P.
6.2
Mounting location
The FMD72 is best suited to level measurement in tanks with pressure overlay, high distillation
columns and other containers with changing ambient temperatures.
The sensor module HP is mounted on the lower measuring connection and the sensor module LP
is mounted above the maximum level. The transmitter can be mounted on pipes or walls with the
mounting bracket.
6.3
Orientation
• Transmitter: Any orientation.
• Sensor modules: The orientation can cause a zero point shift .
This position-dependent zero point shift can be corrected directly at the device via the operating
key, and also in hazardous areas in the case of devices with external operation.
6.4
General installation instructions
Mounting the sensor modules and transmitter is very easy
• The housings of the sensor modules can be rotated up to 380°.
• The transmitter is freely rotatable in the mounting bracket.
The sensor modules and transmitter can be easily aligned when mounted.
Your benefits
• Easy to mount by optimally aligning the housing
• Good, accessible device operation
• Optimum readability of the onsite display (optional)
• Simple pipe installation through optional alignment of the modules
max. 380°
2 mm
A0017506