Description of device parameters
Deltabar FMD71, FMD72
54
Factory setting
• "Pressure" measuring mode: "Sensor temperature HP"
• "Level" measuring mode, Lin. mode "Linear": "Measured pressure LP"
• "Level" measuring mode, Lin. mode "Activate table": "Measured pressure LP"
Display
Depending on the selected measuring mode, the following measured values can be
displayed:
Measured value
Measuring mode
Measured differential pressure
Level, pressure
Corrected pressure
Level, pressure
Measured pressure HP
Level, pressure
Sensor press. HP
Level, pressure
Sensor temperature HP
Level, pressure
Measured pressure LP
Level, pressure
Sensor press. LP
Level, pressure
Sensor temperature LP
Level, pressure
Level before linearization
Level
Tank content
Level
Process density
Level
Temperature of electronics
Level, pressure
4th Value (154)
Navigation
Expert → Communication → HART output → 4th Value (154)
Write permission
No write permissions. Parameter is read only.
Description
The fourth value is displayed.
HART Input Value (155)
Navigation
Expert → Communication → HART Input → HART Input Value (155)
Write permission
No write permissions. Parameter is read only.
Description
Display of the HART input value
HART Input Unit (156)
Navigation
Expert → Communication → HART Input → HART Input Unit (156)
Summary of Contents for Deltabar FMD71