The following options are available to you with an operating tool that is based on FDT
technology (e.g. FieldCare):
• Storage/recovery of configuration data.
• Duplication of device parameters.
• Transfer of all relevant parameters when replacing electronic inserts.
Use the following parameter for this:
Download select. (visible only in FieldCare)
Navigation
Expert → System → Management → Download select.
Write permission
Operators/Service engineers/Expert
Description
Selection of data packages for up/download function in Fieldcare and PDM.
Prerequisite
DIP switch set to "SW" and "Damping" set to "on". If you download using the factory setting
"Configuration copy", all parameters required for a measurement will be downloaded. The
functionality of the "Electronics replace" setting is reserved for Hauser Service and
can be accessed only if the correct device access code is entered.
Options
• Configuration copy: This option overwrites general configuration parameters with the
exception of the serial number, order number, calibration, pos. zero adjust, application
and day information.
• Device replacement: This option overwrites general configuration parameters with the
exception of the serial number, order number, calibration and pos. zero adjust.
• Electronics replace: This option overwrites general configuration parameters.
Factory setting
Configuration copy