Deltabar FMD71, FMD72
Operation with an operating menu
11
Accepting the pressure present
Example: setting position adjustment.
Menu path: Main menu → Setup → Pos. Zero Adjust
Pos. Zero Adjust
007 Operation
1 Cancel
The pressure for position adjustment is present at the device.
Confirm
2
Cancel
Use or to switch to the "Confirm" option. The active option is
highlighted in black.
Confirm
3
Adjustment has
been accepted!
Accept the applied pressure as a position adjustment with the key. The
device confirms the adjustment and goes back to the "Pos. zero adjust"
parameter.
4 Cancel
Exit edit mode for the parameter with .
Confirm
User-definable parameters
Example: setting parameter "Set URV (014)" from 100 mbar (1.5 psi) to 50 mbar (0.75
psi).
Menu path: Setup → Advanced. Setup → Current output → Set URV
Set URV
014
Operation
The onsite display shows the parameter to be changed. The "mbar" unit
is defined in another parameter and cannot be changed here.
1
1 0 0 . 0 0 0
mbar
Press or to get to edit mode. The first digit is highlighted in
black.
2
1
0 0 . 0 0 0
mbar
Use the key to change "1" to "5". Press the key to confirm "5".
Cursor jumps to the next position. Use the key to confirm (second
position).
3
5
0 0 . 0 0 0
mbar
The third digit is highlighted in black and can now be edited.
4
5 0
0
. 0 0 0
mbar
Use the key to change to the "
¿
" symbol. Use to save the new
value and exit edit mode. See next graphic.
5
5 0
¿
. 0 0 0
mbar
The new value for the full scale value is 50.0 mbar (0.75 psi). Exit edit
mode for the parameter with . You can get back to edit mode with
or .
6
5 0 . 0 0 0
mbar
2.5
Operation using Hauser operating program
The FieldCare operating program is an Hauser plant asset management tool
based on FDT technology. With FieldCare, you can configure all Hauser devices as
well as devices from other manufacturers that support the FDT standard.
Hardware and software requirements can be found on the Internet: