Installation
Cleanfit P CPA472
8
Hauser
3
Installation
3.1
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter
has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents. Keep the damaged products until the
matter has been settled.
• Check that the scope of delivery is complete and agrees with your order and the shipping
documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your sales center responsible.
3.2
Installation conditions
3.2.1
Notes on installation
The assembly is designed for installation on tanks and pipes. Suitable nozzles must be available for
this.
!
Note!
• When using standard glass sensors, only installation positions are permitted in which the middle
axis of the assembly lies at an angle of 15° to 75° from the horizontal (see diagram). Otherwise,
there will not be a reliable contact between the inner side of the pH membrane and the inner
terminal leads via the electrolytes. If the angle of the installation position is above 75° from the
horizontal, air bubbles can accumulate in the chamber.
• When using an ISFET Tophit sensor, there are, in principle, no restrictions for the installation
position. An installation angle of 0 to 75° is, however, recommended. An overhead installation is
possible.
a0008528
Fig. 2:
Installation angles for glass sensors (X) and ISFET sensors (Y)
A
Inadmissible installation angle
B
Limited permissible angle (air bubbles can build up in the chamber)
C
Recommended installation angle
90°
75°
15°
0°
0°
15°
75°
A
A
B
C
C
90°
75°
0°
0°
75°
B
C
C
X
Y