Description of parameters
Cerabar S 4 to 20mA HART
34
Hauser
LAST DIAG. CODE (564)
Display
Displays the last messages that occurred and were eliminated.
!
Note!
• Onsite display: you can scroll through the last 15 messages with the
O
or
S
key.
• Digital communication: the last message is displayed.
• Use the RESET ALL ALARMS parameter to delete the messages listed in the LAST
DIAG. CODE parameter.
ACK. ALARM MODE (401)
Options
Switch on acknowledge alarm mode.
See also ACK. ALARM.
Options:
• On
• Off
Factory setting:
Off
ACK. ALARM (500)
Options
Acknowledge alarm.
Prerequisite:
• ACK. ALARM MODE = On
Options:
• Abort
• Confirm
The cause of the alarm must be eliminated, the message must be acknowledged via the
ACK. ALARM parameter and, where applicable, the ALARM DISPL. TIME (
has to have elapsed before the device starts measuring again following an alarm.
See also these Operating Instructions, Section 6.3 "Confirming messages".
Factory setting:
Abort
RESET ALL ALARMS (603)
Options
Use this parameter to reset all the messages of the LAST DIAG. CODE parameter.
Options:
• Abort
• Confirm
Factory setting:
Abort
ERROR No.
Entry
For "Error"-type messages, you can decide whether the device should behave as in the
event of an alarm (A) or as in the event of a warning (W). Enter the corresponding
message number for this parameter.
See also SELECT ALARMTYPE.
See also these Operating Instructions, Section 6.1 "Messages" and Section 6.2
"Response of outputs to errors".
Prerequisite:
• Digital communication
SELECT ALARMTYPE
(595) – Entry
(600) – Selection
For "Error"-type messages, you can decide whether the device should behave as in the
event of an alarm (A) or as in the event of a warning (W).
See also ERROR No.
See also these Operating Instructions, Section 6.2 "Response of outputs to errors".
Options
:
• Alarm (A): output current assumes a defined value.
• Warning (W): device continues measuring
Onsite operation:
1.
Enter the corresponding message number for ERROR No. field.
2.
Select "Alarm" or "Warning" option.
Digital communication
1.
Enter the corresponding message number via the ERROR No. parameter.
2.
Use the SELECT ALARMTYPE parameter to select the "Alarm" or "Warning" option.
Table 15: (GROUP SELECTION
) OPERATING MENU
DIAGNOSTICS
MESSAGES
Parameter name
Description