Fitting the protection cap on the sensor
1. To keep the membrane moist after the sensor has been removed, fill the protection cap
with some electrolyte or clean water.
A0034264
17
Carefully slide protection cap onto the membrane cap.
2. Top part of protection cap is in the open position.
Carefully slide protection cap onto the membrane cap.
3. Secure protection cap by rotating the top part of the protection cap.
A0034494
18
Securing protection cap by rotating the top part
9.2.5
Regenerating the sensor
During measurement, the electrolyte in the sensor is gradually exhausted due to chemical
reactions. The gray-brown silver chloride layer that is applied to the anode at the factory
continues to grow during sensor operation. However, this has no effect on the reaction taking
place at the cathode.