CCS240/241
Technical data
Hauser
33
12.2 Performance characteristics
12.2.1 Response time
CCS240-*
T
90
< 2 minutes
CCS241-*
T
90
< 5 minutes
in applications involving mainly active chlorination
12.2.2 Long-term drift
< 1.5 % per month
12.2.3 Polarization time
Initial commissioning
Recommissioning
CCS240-*
30 min
10 min
CCS241-*
90 min
45 min
12.3 Environment
12.3.1 Ambient temperature range
–5 to 55 °C (20 to 130 °F)
12.3.2 Storage temperature
With electrolyte
5 to 50 °C (40 to 120 °F)
Without electrolyte
–20 to 60 °C (–4 to 140 °F)
12.3.3 Degree of protection
IP68 IP (up to mounting collar Ø 36 mm (1.42"))
12.4 Process
12.4.1 Process temperature
CCS240, CCS241
2 to 45 °C (36 to 113 °F)
12.4.2 Process pressure
max. 1 bar (14.5 psi) absolute, if installed in the Flowfit CCA250 assembly