ATEX II3G
Proline Promag 50
26
Hauser
Terminal assignment and
connection data, power
supply
Terminal assignment and connection data
Terminal assignment and
connection data for signal
circuits with limited energy
!
Note!
The following tables contain values/specifications, which are dependent on the type code (type of measuring
device). Please compare the following type code to the one shown on the nameplate of your measuring device.
A graphic representation of the electrical connections can be found on Page 25.
Terminal assignment of transmitter 53***-***********H
all
transmitters
1 L (+)
2 N (–)
3
Designation
Supply voltage
(as per nameplate)
Protective earth
Functional values
AC: U = 85 to 260 V
or
AC: U = 20 to 55 V, DC: U = 16 to 62 V
Power consumption: 15 VA / 15 W
Caution!
Observe the grounding plans
of the system!
Transmitter
Terminal no. (inputs/outputs)
20 (+)
21 (–)
22 (+)
23 (–)
24 (+)
25 (–)
26 (+)
27 (–)
Assignment
–
–
–
PROFIBUS PA
PA +
PA –
Electric circuit
–
–
–
EEx nA II or
EEx nL IIC
Safety-related
values
–
–
–
U
i
I
i
P
i
L
i
C
i
FNICO
35 V DC
600 mA
8.5 W
≤
10
μ
H
≤
5 nF
Field device
Functional
values
–
–
–
galvanically isolated
U
Bus
I
Bus
9 to 32 V DC
11 mA
IEC 61158-2 (MBP)