ASP station 2000
9 Trouble-shooting
Hauser
97
9
Trouble-shooting
9.1
Trouble-shooting instructions
Trouble-shooting
instructions
Always begin trouble-shooting using the following checklists if faults occur after
commissioning or during operation. Different questions will guide you to the cause of
the error and will suggest appropriate remedial action.
9.2
Process error messages
Process error messages
RPS20A-B3
Composite container PE 30 l
RPS20A-B6
Composite container PE 60 l
RPS20A-VA
Distribution system (tap, tap drive, distribution frame)
50089636
Distribution pan 6x for distribution 12 bottles
50089637
Distribution pan 12x for distribution 24 bottles
RPS20A-PA
PROFIBUS
®
DP slave module
51005197
Glass dosing chamber with bayonet connection
50086064
Air filter
Order code
Accessory
Error message
Cause
Solution
ERROR: RAM
New programme transmitted
Acknowledge the message
ERROR: Clock
Electronics error
Repair by E+H Service
ERROR: EEPROM
New programme transmitted
EEPROM defective
Acknowledge message. Repair
by E+H Service.
ERROR: Conductivity 1
Contact between cond. 1 -
probes due to water or dirt
Clean dosing unit see Section 7
ERROR: Conductivity 2
Contact at cond. 2 - probe due
to water/dirt
Clean dosing unit see Section 7
ERROR: Air manager zero
point
The air manager cannot find the
zero position.
Replace the air manager, if
necessary, repair by E+H
Service.
ERROR: Tap zero point
Tap defective or jammed.
Check the tap, if necessary
replace the distributor unit or
have it repaired by E+H Service.
ERROR: Distributor missing
Distributor not connected to
controller.
Check the connector of the
distributor, if necessary, repair
by E+H Service.
ERROR: 4-20 mA < 3 mA
Analogue signal transmitter
defective, no analogue signal
connected, line break
Check signal current, line and
signal transmitter.
ERROR: Temp. sample
compartment
Temperature in sample
compartment > 60°C,
temperature sensor defective
Check installation conditions
see Section 3.3, repair by E+H
Service.