6 Commissioning
ASP station 2000
90
EHauser
Programme stop options: “Reset” of parallel sampling
Menu structure:
Fig. 42:
Parallel sampling reset menu structure
Define the programme stop “Reset” of parallel sampling as follows:
6.5
Description of device functions
The following table lists and describes all the setting parameters which you can read off
and configure when configuring the device.
no
Parallel sampling is stopped automatically when the last reserve bottle has
been filled.
parall.end
When the last reserve bottle is filled, parallel sampling continues
automatically with the first reserve bottle.
prog. end
When the main programme is completed, parallel sampling continues
automatically at the first reserve bottle.
Parameter
Menu level
Setting options
Factory setting
Info
Info
Set
↵
Info
Act. temp.:
Shows the current temperature in the sample compartment.
Status:
Shows the thermostatic status in the sample compartment.
Act. flow:
Shows the current flow rate value of a connected flowmeter.
Powerfail:
Shows the number and duration of power failures which
occurred.
External stop:
Shows the number and duration of external stops.
Event:
Shows the number and duration of event signals.
Basic settings
Name
Set
↵
Basic settings
↵
Name
Enter the required device name.
ASP 2000
Code
Set
↵
Basic settings
↵
Code
Enter the required user code.
!
Note!
If you no longer know the user code, enter the customer code
6051 to unlock the controller configuration.
0000
Date-time
Set
↵
Basic settings
↵
Date-time
Date:
Enter the current date.
Time:
Enter the local time.
Switch:
Select the mode for switching between summer and
winter time.
Summer time:
Date and value for switching to summer time.
Normal time:
Date and value for switching to winter time.
With automatic
switch between
summer and
winter time