Alternately screw in the gas connection (8 mm) and the white Allen
screw by half-turns. The gas connection must end flush with the upper
edge of the holder. The Allen screw must protrude slightly.
}
To check, slightly unscrew the connection piece and screw it back in.
}
If there is noticeable resistance against the glass body, unscrew the gas
connection and Allen screw by approx. 1 mm and repeat screwing them
in alternately.
}
Mount the inner and outer tube (→ "Cleaning the dismantleable torch"
58).
7.2.3
Maintenance of the one-piece torch
WARNING
Risk of chemical burns from aqua regia
Aqua regia is a 3:1 mixture of concentrated hydrochloric acid and nitric acid. Aqua regia
is highly corrosive and has an oxidizing effect.
¡
Wear protective glasses and clothing when creating and working with aqua regia.
Work under an extraction hood.
¡
Observe all instructions and specifications in the safety data sheets for the basic ma-
terials.
CAUTION
Risk of burns from the hot torch
Plasma is extremely hot. The torch is still very hot after the plasma has been extin-
guished. Contacting these hot surfaces can cause burns.
¡
Wait 5 minutes after extinguishing the plasma. Only touch the torch after this time.
CAUTION
Risk of injury
A risk of injury due to broken glass is present when handling glass parts.
¡
Handle glass parts with extreme caution.