Operation
Biometra TOne
36
Program parameters
Display
(example)
Description
GoTo
Cycles
2
30
Define program loops:
¡
Enter the step number to which the pro-
gram shall return
¡
Define the number of iteration cycles
(max. 999)
ΔT(°C)
10.0
Increase or decrease the block temperature
in each cycle by the temperature increment
(
Temp. incr./decr.:
) of ±0.1 to 20 °C
Δt(s)
120
Extend the holding time in each cycle by the
time increment of 1 to 240 s
ΔR(°C/s)
3.0
Adjust the average heating and cooling rate
for the selected step
The maximum heating and cooling rate de-
pends on the device model.
Degrees
55.0
+1.0
Increase or decrease the block temperature
column by column by the temperature incre-
ment [°C] starting from the annealing tem-
perature [°C]
5.6.5
Editing, adding or deleting steps
Editing steps
There are two options to edit a program step in the software: Directly in the graphical or
spreadsheet view or via an extra page.
ð
Prerequisite: You have opened an existing program or a program template for edit-
ing.
}
Tap on a program step. The selected step is highlighted in blue.
}
Edit the individual parameters for the selected step directly in the spreadsheet or dia-
gram.
}
Define at least one block temperature and holding time for each program step.
}
Tap
GoTo
,
ΔT
,
Δt
or
ΔR
in the graphical view to edit these parameters. Use the data
entry line underneath the diagram.
}
Or: Tap
Edit
and open an extra page for the programming process. Enter all parame-
ters for the step on the page.
}
Use the arrow keys on this page to navigate from step to step. This allows you to pro-
gram several steps one after the other (multi-step programming).
}
Confirm the entered parameters for all steps by tapping the green check mark. There
is no need to confirm the entry for each individual step.