Specification
PlasmaQuant MS product family
134
Altitude
0 to 2000 m
Storage
Temperature: +5 °C to +45 °C, use desiccant
For optimum analytical performance, it is recommended that positive air pressure be
maintained in the laboratory at all times.
Material
Heat and corrosion resistant (recommended V2A steel)
Extraction output
3.0 m
3
/min (110 ft
3
/min) – 4.5 m
3
/min (160 ft
3
/min)
recommended value: 3.5 m
3
/min (125 ft
3
/min)
Adapter using
flexible aluminum or
plastic pipe
Internal pipe diameter
∅
100 mm (4 in)
Length: < 5000 mm (15 ft)
Mass
186 kg
Dimensions
(W x D x H)
660 mm x 589 mm x 1131 mm
Leave a clearance of 30 cm on the right-side of the ICP-MS for opening the interface
door.
15.1.2
Control computer data
Processor
2.8 GHz, or better
Hard Disk Drive
min. 80 GB hard disk drive storage capacity
RAM
4 GB of RAM
Graphic resolution
1280 x 1024 pixels
CD/DVD Drive
16x DVD RW+/- drive
USB Ports
min. 2x USB 2.0 ports
Serial Ports
min 1x
Multimedia
Integrated video, audio, and network support
Operating System
Windows 10 (32- or 64-bit), Windows 7 and 8.1 are supported
15.1.3
Water cooler data
Supply voltage
Frequency
200-240 V AC
50/60 Hz
Cooling Capacity
Minimum of 1.8 kW
Water Temperature
20 °C ± 1 °C (68 °F ± 1.8 °F)
(temperature of the water supplied to the instrument)
Water Pressure
440 ± 40 kPa (64 ± 6 psi)
Pressure at the instrument water inlet
Exhaust unit
Dimensions and weights