multi EA 5100
Sulfur analysis with the S module 5100 (basic, MPO)
69
9.2
Installation
NOTICE
Connecting or disconnecting electrical contacts may damage the sensitive electronic
components of the basic module and of the detection module.
¡
Always connect the modules to power when they are switched off.
}
Place the detection module to the left of the basic module. In the event of a series of
detection modules: place the detection module to the left or right of the others.
}
Connect the supplied power cable to the power cable connection on the rear of the
module and to an earthed socket. Observe the permissible maximum voltage!
}
Connect the detection module with the basic module via the interface:
S-UVF interface on the rear of the detection module
S-UVF interface on the rear of the basic module
}
Connect the measuring gas hose of the basic module to the "sample in" gas inlet on
the rear of the module.
}
Leave the "sample out" outlet unconnected or connect it to the measuring gas inlet of
the next detection module.
}
For detection modules with MPO technology: Install the chemical ozone decomposer
on the rear of the module:
– Fasten the two holding clamps with the supplied screws.
– First press the chemical ozone decomposer into the upper holding clamp and
then into the lower one.
– Connect the hose from the "waste (MPO)" outlet to the ozone decomposer. Do not
pull the hose out of the device!
ü
Connection of the detection module is complete.
Figure 45Chemical ozone decomposer