Maintenance and care
multi EA 5100
154
}
Place the lamp in the proper position when inserting: The pin on the
holder must fit into the groove of the lamp body.
}
Fasten the new lamp with the 2 screws.
}
Plug the connection plug back into the socket all the way.
}
Fasten the side panel again.
ü
The detection module is ready for operation again.
14.12.2 Replacing the chemical ozone decomposer
Only for the S module 5100 MPO
ð
Replace the entire chemical ozone decomposer at least once annually.
ð
Always replace the ozone decomposer when the smell of ozone is noticeable.
}
Undo the hose from the chemical ozone decomposer. Do not pull the hose out of the
device!
}
Pull the ozone decomposer out of the holding clamps.
}
Press the new ozone decomposer into the holding clamps. Reattach the hose.
ü
The detection module is ready for measurement again.
Figure 73Replacing the chemical ozone decomposer