multi EA 5100
Maintenance and care
145
Checking the function of the
detection module
Check the function of the detection module after maintenance via a control measure-
ment.
}
Perform a purge measurement with a solvent, e.g., isooctane.
}
Analyze a standard solution (5 mg/l TN
b
) Compare the curve and area with earlier
measurements.
}
For resuming measuring operations: Determine the daily factor to test calibration. If
the daily factor is outside of the tolerance range, the analysis system must be re-cali-
brated.
14.10.2 Replacing the absorber
ð
Replace the absorber if the analysis baseline has been permanently raised. Replace
the entire absorber (replacement part).
}
Remove the screw connection from the absorber. Do not pull the hose out of the de-
vice!
}
Pull the absorber out of the holding clamps.
}
Press the new absorber into the holding clamps. Reconnect the hose.
ü
The detection module is ready for measurement again.
1
2
3
Figure 69Absorber
1 Hose 6 connection
2 Holding clamp
3 Absorber