multi EA 5100
Troubleshooting
117
Error
Possible cause
Remedy
¡
Sample too viscous
too be drawn in with-
out air pockets
¡
Use the horizontal
operation mode and
dilute the sample, or
dose directly as a
solid.
¡
Inhomogeneous sam-
ple or sample con-
taining particles
¡
Homogenize sample
Low results
¡
Incomplete SO
2
for-
mation due to exces-
sive sulfur content
¡
Reduce sample size/
amount
¡
Dilute sample
¡
Metallic cations in
the sample lead to
sulfur salt formation
¡
Dilute samples, if
possible
Results too high
¡
Excessive halogen
and nitrogen con-
tents disrupt TS de-
termination
¡
Nitrogen: Use MPO
technology
¡
Dilute samples, if
possible
¡
Incomplete sample
incineration (pyroly-
sis products)
¡
Use a suitable com-
bustion mode or
sampler, clean device
before continuing
work
13.13 Device faults on the coulometric sulfur detector
Error
Possible cause
Remedy
Stirring not working
¡
Module not switched
on
¡
Switch on the module
¡
No stirrer rod in the
measuring cell
¡
Insert stirrer rod in
the measuring cell
¡
Stirrer rod faulty
¡
Replace stirrer rod
¡
Magnetic stirrer
faulty
¡
Inform customer ser-
vice department
13.14 Analytical problems during coulometric TS determination
Error
Possible cause
Remedy
Scattering measurements
¡
Sample unsuitable for
vertical operation
mode (droplet forma-
tion)
¡
Use the horizontal
operation mode