Commissioning
5.1
Site requirements
5.1.1
Installation conditions
The following requirements are placed on the climatic conditions in the operating
room of the analyzer:
Temperature range: +10 °C to +35 °C
Max. humidity: 90% at 30 °C
Air pressure: 70 kPa to 106 kPa
As far as possible, the atmosphere in the laboratory should be free of TOC and dust
and free from draft, corrosive and organic vapors and vibration. Smoking is prohibited
in the operating room of the analyzer!
The following requirements are placed on the location of the analyzer:
Do not locate the analyzer directly near a door or window.
Place the analyzer on a heat-resistant and acid-resistant surface.
Do not locate the analyzer near sources of electromagnetic interference.
Avoid direct sunlight and radiation from heaters onto the analyzer. If necessary
provide air conditioning.
Never obstruct the front doors, the side panels and the ventilation slots of the ana-
lyzer with other equipment or furnishings!
Keep a safety distance of at least 5 cm from the back and the right side of the
equipment to other equipment or walls!
5.1.2
Space requirement
The required space depends on the number of components needed for measurement.
Leave adequate space for the PC, monitor, printer and any add-on equipment.
5.1.3
Energy supply
WARNING
The analyzer multi EA 4000 must only be connected to a properly grounded mains
outlet in accordance with the voltage specifications on the type plate!
The multi EA 4000 is operated from the single phase alternating current mains.
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...