Function and setup
multi EA 4000
20
Figure 3 Dust trap at the combustion furnace
The dust trap is arranged easily accessible on the left side of the device behind the
contact guard.
Directly after the dust trap a dust filter has been installed in the gas path. It holds back
particles which the dust trap could not manage to remove from the measuring gas.
Figure 4 Dust filter
The drying tube, filled with magnesium perchlorate as a desiccant, has been installed
in the gas path for drying the measuring gas during C/S determination.
A halide trap (U tube) is has been installed in the basic device to remove interfering
substances from the measuring gas and to protect the detectors and the flow regula-
tor. The U tube is filled with special copper wool and brass wool. The filling of the hal-
ide trap has to be renewed at the latest once half of the copper wool or the brass wool
is discolored.
Dust filter
Drying tube and
halide trap for
C/S determination
Summary of Contents for Analytik Jena multi EA 4000
Page 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Page 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Page 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Page 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Page 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Page 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Page 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Page 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...