Operation with IPextract
InnuPure C16
touch
48
6.6
Checking and executing individual device functions with tools
Individual device functions can be executed and checked using special programs
(tools). These tools allow, for example, for the filter tips to be emptied and ejected af-
ter abortion of an extraction protocol.
The following tools have been integrated into the program:
Icon
Tool
Function
T
AKE
T
IPS
Take filter tips from the tip block.
E
MPTY
T
IPS
Emergency emptying of strips and plates after protocol
abortion into any cavity to discharge fluid from the tips in a
controlled manner. The sample tray must be populated with
corresponding strips and plates.
H
EATING TEST
Test of the heating function
In this test, a defined temperature is set and held for a pe-
riod. The setting of the temperature and the temperature
constant during the holding time are documented. The log
file of the test will be saved in the local user folder of the
tablet.
I
NITIALIZATION
Initialization of the InnuPure C16
touch
after protocol abor-
tion or after an error occurred
R
EMOVE
T
IPS
Ejection of the filter tips into the 3rd cavity after a protocol
abortion. The sample tray must be populated with corre-
sponding strips and plates.
M
AGNET
T
EST
Check positioning of the magnet unit
T
RAY
O
UT
Ejection of the sample tray from the device
D
ECONTAMINATION
With installed decontamination light only:
Carry out decontamination with UV light
L
IGHT
Switching the interior light on/off separately
1.
Select the menu item O
PTIONS
T
OOLS
from the H
OME
page.
2.
Tap on the tool button.
3.
Follow further instructions on the tablet.
Integrated tools
Using the tools