CyBio Well vario
Operation
the middle of its companion well.
Fig. 87:
A1 well target positions for 96/384/1536 well microplate
1 A1 well target position for 96 well microplate
2 A1 well target position for 384 well microplate
3 A1 well target position for 1536 well microplate
The diagram view illustrates that the tips will come down onto the edge of wells in the case
of a faulty plate type setting.
Example:
If a 1536 well microplate were to be processed with a 96 tip magazine, the tip would come
down in point 1 of figure 56.
Solution:
Set 1536 for plate type in
[Configuration]
=>
[Wells]
menu. Perform manual motion to
A1 well position for 1536 well microplates before dispensing (point 3 in figure 56).
XY
This menu item is intended for storing internal XY correction values for lifter motion. In the
event of necessary lifter replacement (according to manufacturer specification provided on
lifter underside), you can update your lifter settings here.
Fig. 88:
Display – XY correction values
The first entries in a line indicate the direction. The coordinates for zero point (nn) corre-
spond to those of a 96 well microplate.
The path, however, represents the distance to a 384 well or 1536 well in the correspond-
ing coordinate direction (refer to details on underside of lifter).
On selection of a coordinate, the lifter moves to the assigned position and can thus be eas-
ily inspected for correct adjustment. A microplate of transparent material is recommended
for lifter adjustment.
Stage speed
A stage speed setting will be valid for lifter motion, carriage motion and rotary stage
motion. You can enter an end value for motion speed here. Software settings can be made
for end speed, or starting speed and end speed or for the motion characteristic.
1
2
3
CAUTION
With excessively high settings for carriage motion or rotary stage motion, there may be
overflow of liquid from the reagent troughs. Select a reasonably reduced speed value if
necessary.
TIP
In manual mode, lifter motion is limited to a maximum speed of 17 mm/s (60 rpm).
This is necessary to enable a more precise height setting. In program-controlled (com-
puter-controlled) mode, the limiting value is around 40 mm/s (140 rpm).
Summary of Contents for Analytik Jena CyBio Well vario
Page 1: ...User Manual CyBio Well vario ...
Page 3: ...Rev B 2022 03 III CyBio Well vario ...
Page 5: ...Rev B 2022 03 V CyBio Well vario ...
Page 6: ...VI Rev B 2022 03 CyBio Well vario ...
Page 7: ...Rev B 2022 03 VII CyBio Well vario ...
Page 8: ...VIII Rev B 2022 03 CyBio Well vario ...
Page 12: ...XII Rev B 2022 03 CyBio Well vario ...
Page 16: ...XVI Rev B 2022 03 CyBio Well vario ...
Page 26: ...10 Rev B 2022 03 Technical Specifications CyBio Well vario ...
Page 62: ...46 Rev B 2022 03 Technical Description CyBio Well vario ...
Page 118: ...102 Rev B 2022 03 Troubleshooting CyBio Well vario ...
Page 126: ...110 Rev B 2022 03 Decommissioning CyBio Well vario ...
Page 130: ...114 Rev B 2022 03 Accessories Spare Parts CyBio Well vario ...
Page 132: ...116 Rev B 2022 03 Waste Disposal CyBio Well vario ...
Page 142: ...126 Rev B 2022 03 CyBio Well vario ...