Automatic Boat Drive
Maintenance and care
31
}
Re-install the ABD at the basic module (→ "Installing the ABD at the
combustion tube"
25).
ü
The system is now ready for operation again.
7.8
Cleaning or replacing the guide tube
NOTICE
If a sampler has been fitted to the ABD, this must be removed prior to removing the
guide tube (see user manual Multi Matrix Sampler).
If the guide tube needs cleaning or replacing, proceed as follows:
}
Switch the ABD off from the equipment switch.
}
Disconnect the ABD from the basic module (→ "Disconnecting the ABD
from the combustion tube"
27).
}
Remove the hook from the guide tube (→ "Cleaning and replacing the
hook at the ABD"
30).
}
Unscrew the argon hose (arrow) at the block of the gas supply.
}
Using the face spanner, loosen the clamping ring at the right side of the
sample sluice.
}
Loosen the clamping ring at the block.