ENDEVCO Isomin 25B Instruction Manual Download Page 1

INSTRUCTION  MANUAL 

For Model 25B

,  

Isomin™ Accelerometer 

IM25B, Revision A

 

IMPORTANT:  To meet all published specifications, the constant current source used to 

power this accelerometer must be adjusted to 4.0 ±0.5 mA. 

The Endevco Model 25B is a microminiature, piezoelectric accelerometer with integral electronics. Because of 

its extremely small size and physical construction, certain special precautions must be observed when 

mounting and removing the accelerometer to avoid damaging the unit. 

Accessories, Adhesive, Solvent and Tools 

For use at room temperature: 

* Petro wax or bees wax

For use up to 60°C: 

* All-temperature glue stick and hot-melt glue gun

* Solvent:  Follow the recommendations from glue stick manufacturers

For use up to 125°C: 

* Adhesive: Cyanoacrylate (Loctite Super Bonder 430 Adhesive is recommended)

* Solvent: Nitromethane base (Loctite X-NMS Solvent 768 is recommended)

Alternatives: Acetone, Chlorothene (Trichloroethane) or equivalent hydrocarbon solvent

* Cotton swabs or lint-free tissue

* Toothpicks or equivalent applicators

Endevco P/N 31849 Adhesive Kit contains most of the required items for mounting transducer.  Contact 

the factory for details. 

CAUTION:

  

Observe precautions recommended by manufacturer of adhesive or solvent. 

WARNING:

   When Cyanoacrylate is used, do not break cement bond by striking accelerometer or by 

using any tool that makes a metal-to-metal contact with the accelerometer.  Always 

break the bond with the supplied tool after  cement has  been softened, by twisting or 

torquing the accelerometer, stressing the bond in shear. 