ENCARDIO RITE ESM-40S User Manual Download Page 23

Users’ Manual 

ESM-40S High sensitivity settlement monitoring system 

 

 

Page

 | 21

 

8  CONNECTING EPS-40S TO MODBUS DATALOGGER 

8.1  Introduction 

Encardio-rite model ESM-40S settlement monitoring system use the industry standard Modbus Remote 
Terminal Unit (RTU) protocol to communicate with dataloggers. As the name suggests, Modbus was 
designed to work on what is known as a bus network, meaning that every device receives every message 
which passes across the network.  
Model ESM-40S sensors use the RS-485 electrical interface to communicate over long distance cables up 
to  1.2 km.  Any suitable Modbus datalogger with either GSM/GPRS or RF transmission facility can be 
connected for datalogging and wireless data transmission. 
Encardio-rite model ESDL-30 datalogger can also be used with an additional Modbus card, if specifically 
ordered. However, maximum limit of Modbus seepage sensors that can be connected to this ESDL-30 
datalogger version is seven. 

8.2  Connection details for ESDL-30 datalogger 

ESDL-30 datalogger has a Modbus (RS-485) port also besides the three SDI-12 ports (channels). Sensors 
with Modbus (RS-485) output can be connected to the datalogger as shown in figure below. Battery should 
be placed in the datalogger after the connections have been successfully done. 

8.3  Wiring details for any other Modbus datalogger 

 

Figure 8-2 Wiring details to connect ESM-40S sensor with Modbus output to suitable datalogger 

 

  

  

Modbus 

interface 

Modbus Sensor 

Addr: 0x01 

 

  

Red: +12 V

 

Black: GND

 

Green: D+

 

White: D-

 

  

  

Modbus Sensor 

Addr: 0x02 

 

Modbus Sensor 

Addr: 0x03 

 

Bus cable 

Figure 8-1 Connection of Modbus (RS-485) 4-core cable bus (from settlement sensor 

chain) to ESDL-30 datalogger

 

Summary of Contents for ESM-40S

Page 1: ...a P 91 522 2661039 42 Email geotech encardio com www encardio com International India Bhutan Bahrain Qatar UAE Morocco Greece Spain UK USA India Lucknow Delhi Kolkata Mumbai Chennai Bangalore Hyderaba...

Page 2: ...allation of reservoir moisture trap 9 4 4 Installation of pressure gage and valve assembly 9 4 5 Installation of datalogger 10 4 6 Connecting the liquid line 10 4 7 Connecting the air vent line 11 4 8...

Page 3: ...ied in two options SDI 12 serial interface or Modbus RS 485 output The sensors are user friendly and convenient to configure Encardio rite model ESDL 30 datalogger can be used for remote automatic dat...

Page 4: ...allation work The intricacies involved in installation are such that even if a single essential but apparently minor requirement is ignored or overlooked the most reliable of instruments will be rende...

Page 5: ...nt and changes in barometric pressure 2 2 Operating principle The settlement sensors are mounted at locations where settlement is to be monitored These are connected in series by a fluid filled tube a...

Page 6: ...30 compact digital datalogger for remote online monitoring With ESDL 30 datalogger and its configuration software it is quite easy and fast to configure the whole system 2 3 ESM 40S 1 Settlement sens...

Page 7: ...he reference sensor and another vent tube from the reservoir 2 6 ESM 40S 4 Fluid and air vent tubes 8 mm od tubes are used as fluid and air vent tubes Air vent tube is required to compensate the barom...

Page 8: ...ter if the tubes are free of air bubbles The fluid mixture can be arranged locally by user or can be procured from Encardio rite separately The quantity required will depend on the total length of flu...

Page 9: ...re required for installation of the settlement monitoring system Impact or percussion electric drill with 8 mm and 6 mm drill bit for drilling in concrete or masonry for fixing fasteners Cable Cutter...

Page 10: ...all the level sensors of that particular measurement group can be fixed at the same elevation at the chosen locations The sensors are supplied with the mounting plate attached Mounting plate is design...

Page 11: ...e little above the top of reservoir figure 4 6 Drill 6 mm dia x 50 mm deep holes Fix the moisture trap using the four fasteners provided in supply Connect the moisture trap to the reservoir with the d...

Page 12: ...ded Expandable anchor HPS 1 6 15x40 Holes c c 204x82 ESDL 30 Datalogger Wall Cable from SDI 12 sensor Computer 12V power supply Figure 4 4 Mounting details for ESDL 30 datalogger installation on wall...

Page 13: ...rvoir fluid line and sensors Once the complete system is installed fluid will be filled into the reservoir fluid line reference settlement sensor and all the installed settlement sensors It is recomme...

Page 14: ...sensors 11 Repeat the cycle steps 1 to 10 till all the sensors are filled with fluid 12 Open air bleed screw of each sensor to remove trapped air Open it for 3 4 seconds and close it immediately CAUTI...

Page 15: ...with a unique address which is represented by a single ASCII character This communication is achieved by digital communication along a single serial line The digital addressing system allows an SDI re...

Page 16: ...he settlement sensor with Modbus RS 485 interface output requires 4 core bus cable Connection details to male and female connectors is shown in figure below 5 pin male connector Wiring diagram 5 2 4 1...

Page 17: ...tput Data from the datalogger can be transmitted remotely to a central cloud server via cellular GSM GPRS network The datalogger is very easy to configure especially for ESM 40S sensors It provides da...

Page 18: ...ng system Page 16 Connection detail is shown in figure below Battery should be placed in the datalogger after the connections have been successfully done Figure 6 2 Connection of SDI 12 three core cab...

Page 19: ...nfiguration for model ESM 40S sensor with SDI 12 output is given below 1 Double click the ESDL 30 datalogger software icon on the Desktop The main screen of the application appears as displayed in fig...

Page 20: ...ite for first time Figure 7 3 Select site to open 5 Click on Config Sensors tab and the select Smart Config Sensor as shown in figure below Figure 7 4 6 The screen as shown in figure 7 5 will appear S...

Page 21: ...y settlement monitoring system Page 19 Figure 7 5 Figure 7 6 Smart config sensor screen 7 Screen as shown in figure 7 7 will appear Press Read button The application will fetch the information stored...

Page 22: ...ensitivity settlement monitoring system Page 20 8 Press Monitor Sensor button The screen as shown in figure 7 9 will appear showing the values of settlement reading Parameter 1 and temperature reading...

Page 23: ...te model ESDL 30 datalogger can also be used with an additional Modbus card if specifically ordered However maximum limit of Modbus seepage sensors that can be connected to this ESDL 30 datalogger ver...

Page 24: ...example the most recent sensor reading is stored in a table This reading is presented in different formats in different sections of the table The register location and size of these variables is deta...

Page 25: ...es 00 00 Register value Ch 1 Status 16 bit unsigned 00 Number of registers Hi bytes 3B 21 3E 1B Register value Parameter 32 bit Float 0 002460 06 Number of registers Lo bytes 00 00 Register value Ch 2...

Page 26: ...ntenna a microcontroller sensor module that reads the sensor a wireless communication module that transmits the digital data to the gateway and a processor that controls the two modules It is a highly...

Page 27: ...S 485 output option The sensor with Modbus RS 485 output will be connected to the RS485 port center one of the digital node as shown in figure below Figure 8 3 Connection of digital sensor with Modbus...

Page 28: ...tion Table Marking on Standard Tape mm Reading Observed by Automatic level sensor mm Error mm 0 0 00 0 10 200 00 200 10 0 25 400 00 400 25 0 25 600 00 600 25 0 15 800 00 800 15 0 25 1000 00 1000 25 0...

Page 29: ...d the most reliable of instruments will be rendered useless The warranty is limited to as stated herein Encardio rite is not responsible for any consequential damages experienced by the user There are...

Reviews: