Measurements
7.1
Calibrating pH sensors
7.1.1
Calibration overview
The calibration menu allows you to calibrate the pH and temperature response of the
sensor. Calibrations can be performed using the manual or the auto-calibration option. In
both cases, the transmitter does a two-point calibration for pH and one-point
standardization against a reference thermometer for temperature. In auto-calibration, the
transmitter automatically stores the temperature-corrected calibration data. In manual
calibration, you have to enter buffer values. The values are only registered when the
readings are stable (automatically determined by the transmitter). All calibration
procedures are guided by prompts on the display.
7.1.2
Calibration standards (buffer solutions)
•
Calibrations are critical to make accurate pH measurements.
•
It is recommended that the value of one of the pH buffer solutions is lower than the
pH of the process stream and the value of the other pH buffer solution is higher than
the pH of the process stream.
•
For best results, make sure that the temperature of the solution is the same
temperature as that of the sensor. Allow the entire measurement cell, sensor, and
solution to reach relatively constant temperature.
•
Using buffers at high temperatures can cause evaporation, which will change the
concentration of the buffer. The change in concentration causes a change in the pH,
thus resulting in an inaccurate calibration.
•
Buffers have limited shelf lives. Do not use a buffer if the expiration date has passed.
Store buffers at controlled room temperatures.
•
Used buffer solutions should not be reused. Buffers should also be protected from
excessive exposure to air. Exposure to air can cause changes in the pH of the buffer
solution.
•
Always rinse the sensor with DI water and remove excess water by dabbing with a
clean tissue before placing it in a buffer. Only dab the sensor and do not wipe it.
Wiping the sensor builds static charge which alters the reading.
7.1.3
Auto-calibration procedure
Prerequisites
Ensure that auto-calibration is turned ON and the appropriate buffers are used.
Measurements
Instruction Manual
41
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...