c. Press
ENTER
.
The display shows the sensor current.
d. Note the units
µA is microamps; na is nanoamps.
4.
Change the decade box resistance and verify that the correct current is shown.
Calculate the current from the equation:
The voltage of a fresh 1.5 volt battery is about 1.6 volt (1600 mV).
8.4.11
Simulating inputs - pH
General information about simulating pH inputs
This section describes how to simulate a pH input into the transmitter. To simulate a pH
measurement, connect a standard millivolt source to the transmitter. If the transmitter is
working properly, it will accurately measure the input voltage and convert it to pH.
Although the general procedure is the same, the wiring details depend on whether the
preamplifier is in the sensor, a junction box, or the transmitter.
Preamplifier in transmitter
To simulate a pH input when the preamplifier is in the transmitter, complete the following
steps.
1.
Turn off automatic temperature correction and set the manual temperature to 25 °C
(77 °F).
2.
Disconnect the pH sensor. Also, disconnect the chlorine sensor anode lead. Connect
a jumper wire between the pH IN and REF IN terminals.
3.
Confirm that the transmitter is reading the correct mV value.
a. With the main display showing, press
DIAG
.
b. Press
NEXT
until the display shows PH.
c. Press
ENTER
.
The display shows InpUt followed by a number. The number is the raw input
signal in millivolts. The measured voltage should be 0 mV.
4.
Confirm that the transmitter is reading the correct pH value.
a. Go to the main display.
b. Press
3
or
5
.
Diagnostics and troubleshooting
114
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...