
Readings drift.
1.
Is the membrane clean? For the sensor to work properly, oxygen must diffuse freely
through the membrane. A coating on the membrane will interfere with the passage
of oxygen, resulting in a slow response.
2.
Is the sensor in direct sunlight? If the sensor is in direct sunlight during air
calibration, readings will drift as the sensor warms up. Because the temperature
reading lags behind the true temperature of the membrane, calibrating the sensor
in direct sunlight may introduce an error.
3.
Is the sample flow within the recommended range? Gradual loss of flow will cause
downward drift.
4.
Is the sensor new or has it been recently serviced? New or rebuilt sensors may
require several hours to stabilize.
Sensor not responding to oxygen changes
1.
If readings are being compared with a portable laboratory instrument, verify that
the laboratory instrument is working.
2.
Is the membrane clean? Clean the membrane and replace it as necessary. Check
that the holes at the base of the cathode stem are open. Use a straightened paper
clip to clear blockages. Replace the electrolyte solution.
3.
Replace the sensor.
Oxygen readings too low
1.
Low readings can be caused by zeroing the sensor before the residual current has
reached a stable minimum value. Residual current is the current the sensor
generates even when no oxygen is in the sample. Because the residual current is
subtracted from subsequent measured currents, zeroing before the current is a
minimum can lead to low results.
Eaxample: The true residual (zero) current for a 499ADO sensor is 0.05 µA, and the
sensitivity based on calibration in water-saturated air is 2.35 µA/ppm. Assume the
measured current is 2.00 µA. the true concentration is (2.00 - 0.05)/2.35 or 0.83
ppm. If the sensor was zeroed prematurely when the current was 0.2 µA, the
measured concentration will be (2.00 - 0.02)/2.35 or 0.77 ppm. The error is 7.2%.
Suppose the measured current is 5.00 µA. The true concentration is 2.11 ppm, and
the measured concentration is 2.05 ppm. The error is now 3.3%. The absolute
difference between the readings remains the same, 0.06 ppm.
2.
Sensor response depends on flow. If the flow is too low, readings will be low and flow
sensitive. Verify that the flow past the sensor equals or exceeds the minimum value.
See the sensor instruction manual for recommended flows. If the sensor is in an
aeration basin, move the sensor to an area where the flow or agitation is greater.
Diagnostics and troubleshooting
102
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...