
There is a high probability that the measurement is in error. A fault alerts you that a
disabling condition exists. If a fault message is showing, all measurements should be
regarded as erroneous.
When a warning condition exists:
1.
The main display remains stable. It does not flash.
2.
A warning message appears alternately with the temperature and output readings
in the second line of the display. See
for an explanation of the warning
messages and suggested ways of correcting the problem.
When a fault exists:
1.
The main display flashes.
2.
The words FAULT and HOLD appear in the main display.
3.
A fault message appears alternately with the temperature and output readings in
the second line of the display. See
for an explanation of the fault
messages and suggested ways of correcting the problem.
4.
The output current will remain at the present value or go to the programmed fault
value.
5.
If the transmitter is in HOLD when the fault occurs, the output remains at the
programmed hold value. To alert you that a fault exists, the word FAULT appears in
the main display, and the display flashes. A fault or diagnostic message also appears.
6.
If the transmitter is simulating an output current when the fault occurs, the
transmitter continues to generate the simulated current. To alert you that a fault
exists, the word FAULT appears in the display, and the display flashes.
8.4.2
Troubleshooting fault or warning messages
The following table explains the fault or warning messages displayed on the 5081-A
transmitter.
Message
Explanation
See section
OuEr rAnGe
Over range, measurement exceeds display lim-
it
AMP FAIL
Amperometric sensor failure, sensor current is
too high
bAd SEnSor
Bad sensor, sensor current is a large negative
number
0 too biG
Zero current is too large, sensor was zeroed
while current exceeded 100 nA
CAL Error
Calibration error, sensitivity (nA/ppm) is too
high or too low
nEEd 0 CAL
Sensor needs re-zeroing, reading is too nega-
tive
Diagnostics and troubleshooting
94
Rosemount 5081
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...