FAct FAIL procedure
FACt FAIL
appears if the transmitter factory calibration message has been triggered. A stray
noise spike can cause this message to appear. If the pH reading seems acceptable, reset
the calibration flag.
1.
Enter the factory calibration menu by pressing
2
on the IRC ten times.
The display does not change.
2.
Immediately press
3
.
FActorYCAL
appears in the display.
3.
Press
NEXT
.
rEPAir
appears in the display.
4.
Press
NEXT
.
ConFIG
appears in the display.
5.
Press
NEXT
.
rESEt
appears in the display.
6.
Press
ENTER
.
rESEtCFG
appears in the display.
7.
Press
ENTER
.
rESEt
appears again.
8.
Press
NEXT
.
FActorYCAL
reappears.
9.
Press
ENTER
.
FactOn
appears in the display.
10.
Press
3
.
FactOFF
appears.
11.
Press
ENTER
to store the settings.
12.
Press
EXIT
repeatedly until the main display reappears.
If the message does not clear or problems persist, the electronics have failed. Replace the
electronic board stack.
Diagnostics and troubleshooting
Instruction Manual
91
Summary of Contents for Rosemount5081
Page 4: ......
Page 10: ...Contents vi Rosemount 5081 ...
Page 12: ...Startup procedure 2 Rosemount 5081 ...
Page 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Page 24: ...Installation 14 Rosemount 5081 ...
Page 28: ...Wiring 18 Rosemount 5081 ...
Page 50: ...Programming basics 40 Rosemount 5081 ...
Page 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Page 136: ...Digital communications 126 Rosemount 5081 ...
Page 140: ...Engineering drawings 130 Rosemount 5081 ...
Page 141: ...Engineering drawings Instruction Manual 131 ...
Page 142: ...Engineering drawings 132 Rosemount 5081 ...
Page 143: ...Engineering drawings Instruction Manual 133 ...
Page 144: ...Engineering drawings 134 Rosemount 5081 ...
Page 145: ...Engineering drawings Instruction Manual 135 ...
Page 146: ...Engineering drawings 136 Rosemount 5081 ...
Page 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...