Measuring Principles
3.4 Trace Moisture Measurement
3.4.2 Accompanying Gases
Component
Maximum Permitted
Concentration / ppm
Maximum Permitted Dew-
point / °C
Acetylene (Ethyne)
C
2
H
2
1
-20
Ammonia
NH
3
1000
-20
Aromatic alcohols
no limit
no limit
Benzene
C
6
H
6
no limit
no limit
Bromine
Br
2
no limit
-20
Carbon dioxide
CO
2
no limit
no limit
Carbon disulphide
CS
2
no limit
no limit
Carbon monoxide
CO
no limit
no limit
Carbon tetrachloride
CCl
4
no limit
no limit
Carbon tetrafluoride
CF
4
no limit
-20
Chlorine
Cl
2
not permitted
Dichlorodifluoromethane
CCl
2
F
2
no limit
-20
Ethane
C
2
H
6
no limit
no limit
Ethylene (Ethene)
C
2
H
4
no limit
no limit
Ethylene oxide
CH
4
O
not permitted
Exhaust gases
no limit
no limit
Fluorine
F
2
10
-20
Glycol (Ethane-1,2-diol)
HOCH
2
CH
2
OH
no limit
no limit
Halogenated hydrocarbons
Consult with Emerson
Hydrobromic acid
HBr
not permitted
Hydrochloric acid
HCl
not permitted
2
Hydrofluoric acid
HF
500
-20
Hydrogen peroxide
H
2
O
2
not permitted
Hydrogen sulphide
H
2
S
no limit
3
no limit
Mercury
Hg
not permitted
4
Methane
CH
4
no limit
no limit
Methanoic acid
HCOOH
not permitted
Methanol
CH
3
OH
5
no limit
Methylethyl glycol
C
4
H
11
O
no limit
no limit
Natural gas
no limit
no limit
Nitric acid
HNO
3
10
-20
Nitrogen dioxide
NO
2
no limit
-20
Nitrous oxide
N
2
O
no limit
-20
Tab. 3-7: Limitations on Gases
Several gases may affect the sensor, so consider the following limits:
(I)
Summary of Contents for Rosemount X-Stream Series
Page 535: ......