Emerson Process Management GmbH & Co. OHG
8-10
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
Message
Status level
Description
Recommended Actions
Range U
nderflow
Off-spec
Gas concentration is out of
measurement range and the-
refore linearization curve does
not apply (measuring results
are not reliable).
•
Select lower range (
polyno-
mial linearization mode only
)
•
Adjust gas concentration to
be within range
Secondary Sensor Signal
Simulation
Function check
Any secondary sensor’s
signal is simulated for service
purposes
•
Restart device.
•
Ask service personnel to
deactivate simulation.
Sensor ADC
Failure
Input voltage applied to an
internal DC signal input (DC
1...5) too high
•
Adjust sensors output volta-
ge to be within 0 .. 5 V limit
•
Replace sensor
Input voltage applied to an
internal AC signal input (WS
1...4)too high
•
Adjust sensors output volta-
ge to be within ± 6 V limit
•
Replace sensor
Sensor Chopper
Failure
Internal failure bit of electro
-
nics board XSP is set
•
Switch analyzer off and on
again
•
Check red LED on chopper
board UCC
•
Replace chopper
Sensor Communication
Timeout
Failure
The serial communication bet
-
ween the main controller and
the sensor interface has timed
out. The reason is unknown.
Check both boards, and pro
-
per connections
Sensor Detector
Failure
XSP's failure bit was set
•
Switch off / on the analyzer
•
Check if VVS signal is proper
•
Replace detector
Sensor Flow
Failure
The flow sensor is not working
properly
Check the sensors function,
and if need be, replace the
sensor.
Sensor Pressure
Off-spec
The pressure measurement is
not working properly for com-
pensation purposes
Configure pressure to be
within limits
Sensor Signal Simulation
Function check
The primary sensor signal is
simulated for service purpo-
ses
•
Restart device.
•
Ask service personnel to
deactivate simulation.
8.2.2 Channel Related Messages