Software Menus
6.2.3 Setup Menu
6.2.3.4.3
Setup Digital Inputs
Setup..
In-/Outputs..
Digital Inputs..
Digital Inputs (X4.1) 1of3
Input1 Node:
System
Input1 Function:
Span All
Input1 Edge:
Rising
Input2 Node:
Ch1
Input2 Function:
Range1
Input2 Edge:
Trailing
If your instrument features optional digital
I/O boards, this menu appears, enabling to
configure the digital inputs.
Menu pages 1...3 (titled "Digital inputs (X4.1)")
enable to configure inputs 1...7 on the first ex
-
pansion board ('X4.1' in the menu title refers
to the instrument´s I/O connector X4.1).
For each input 1...7 specify within the "
Node
"
line the signal source.
Available options:
System
,
Ch1 ... Ch5
(depending on the number of channels in
-
stalled).
If any one of
Ch1 ... Ch5
is selected, only
signals, valid for the selected channel are
selectable.
If
System
is selected, any system signal is
selectable.
Once the "
Node
" is specified, for each input
•
select the "
Function
" of that input (depen-
ding on the node, the list of available signals
varies;
next page)
• select, how the input is to be triggered: by
Rising
edge, or
Trailing
edge.
Input3 Node:
System
Input3 Function:
Zero All
Input3 Edge:
Raising
Input4 Node:
System
Input4 Function:
ExtFail
Input4 Edge:
Up
Input5 Node:
Ch1
Input5 Function:
Range1
Input5 Edge:
Down
Digital inputs (X4.1) 3of3
Input6 Node:
System
Input6 Function:
Pump1
Input6 Edge:
Trailing
Input7 Node:
Ch3
Input7 Function:
Range1
Input7
Edge:
Rising