Emerson Process Management GmbH & Co. OHG
6-26
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
6.2.3 Setup Menu
6.2.3.1.2.1.1
Component Unit Setup Menu
Setup..
Display..
Measurements..
Component..
Unit..
Ch1
Component Unit
Unit:
ppm
Unit Text:
ppm
Span Gas Unit:
ppm
Span Unit Text:
ppm
Custom Factor:
1.000000
Custom
O
ffset:
0.000000
Span Gas
50000.000 ppm
Concentration
0.000 ppm
Within this menu, configure the component´s
unit to be used for measurement results.
Set the text for the measuring unit of the com-
ponent, to be shown in the measuring screen:
each character must be set separately.
1)
Set the text for the measuring unit of the span
gas: each character must be set separately.
1)
Select the measuring unit for the compo-
nent.
Available options:
ppm, ppb, Vol%, Cu-
stom
Select the measuring unit for the span gas.
Available options:
ppm, ppb, Vol%, Cu-
stom
Note!
Zero gas unit and zero gas unit text are al-
ways as configured for the component (first
2 menu lines)!
Gas concentrations are internally calculated
as ppm. To use other units, the corresponding
factor must always be specified, e.g. 0.0001
to calculate from ppm to %.
2)
If necessary, enter an offset here, to be added
to the measured value.
2)
The last 2 lines show how the settings affect
the display of measurements.
Note!
Texts for units, and values for
"Custom Fac-
tor"
and "
Custom Offset"
are not checked for
plausibility. Any arbitrary value can be set.
1)
Configuring individual text strings is
permitted only, if "
Unit
" or "
Span gas unit
"
is set to
Custom
.
2)
"
Custom factor
" and "
Custom offset
" are
visible and editable only, if "
Unit
" or "
Span
gas unit
" is set to
Custom
. For all other
options, conversion factors and offsets are
pre-defined.