9
Rosemount Series 58C Platinum RTD
9.1
Specifications
9.1.1
Performance specifications
Temperature range
–50 to 200 °C (–58 to 392 °F)
Maximum hysteresis
±0.09 percent of operating temperature range
Stability
±0.035 percent maximum ice-point resistance shift following 1,000 hours at
maximum specified temperature 200 °C (392 °F)
Insulation resistance
500 x10
6
ohms minimum insulation resistance when measured at 50 Vdc at
room temperature
9.1.2
Environmental specifications
Humidity limits
No permanent rear seal is installed.
Quality assurance
Each sensor is subjected to a resistance accuracy test at 0 °C (32 °F) and an
insulation resistance test.
9.1.3
Physical specifications
Sheath material
316 SST
Lead wires
PTFE-insulated, nickel-coated, 24-gauge stranded copper wire
9.1.4
Functional specifications
Power
Overvoltage Category I
Environmental
Pollution Degree 4