6.4.4
Procedure - manual calibration
Complete the following steps to manually calibrate the pH sensor.
Procedure
1. Obtain two buffer solutions.
Ideally, the buffer pH values should bracket the range of pH to be measured.
2. Remove the sensor from the flow cell. If the process and buffer temperatures are
appreciably different, place the sensor in a container of tap water at the buffer
temperature.
Do not start the calibration until the sensor has reached the buffer temperature.
3. Press
MENU
.
The main
Menu
screen appears. The cursor is on Calibrate.
4. Press
ENTER
.
5. Choose the sensor you wish to calibrate.
Sensor 1 is the chlorine sensor. Sensor 2 is the pH sensor.
6. Choose pH.
7. Choose Buffer Cal.
8. Choose Manual.
9. Choose Buffer 1.
10. Rinse the sensor with water and place it in the first buffer. Be sure the glass bulb and
reference junction are completely submerged. Swirl the sensor.
11. Watch the pH reading for sensor 2 (S2) at the top of the screen. Once the reading is
stable, enter the pH value of the buffer at the buffer temperature and press
ENTER
.
The display returns to the screen shown in step 9.
12. Choose Buffer 2.
13. Remove the sensor from the first buffer.
14. Rinse with water and place it in the second buffer. Be sure the glass bulb and
reference junction are completely submerged. Swirl the sensor.
15. Press
ENTER
.
16. Watch the pH reading for sensor 2 (S2) at the top of the screen. Once the reading is
stable, enter the pH value of the buffer at the buffer temperature and press
ENTER
.
If the calibration is successful, the screen below is displayed for five seconds. The
display then returns to the screen in step 7. If the calibration is not successful, the
existing calibration data is not changed. A screen appears identifying the error (high
slope, low slope, or offset error). For troubleshooting, see
17. To return to the main display, press
MENU
and then
EXIT
.
Calibrate
Reference Manual
May 2019
00809-0100-3412
72
Emerson.com/Rosemount
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...