c) If you have a Rosemount FCL-02, choose either Live/Continuous or Manual
for Free Cl Correct (free chlorine correction). Live/Continuous means the
transmitter will use the pH measured on the second channel to continuously
correct the chlorine reading for changes in the sample pH. Manual means the
transmitter will use a fixed pH value entered by you to convert the raw
chlorine signal to a ppm reading.
d) If you have a Rosemount FCL-01, Free Cl Correct (free chlorine correction)
will not appear. Instead, enter the desired pH correction value for Manual pH.
The screen summarizes the present configuration for sensor 2 (pH). There are six
items: Measure, Preamp, Sol'n Temp Corr, Resolution, Filter, and Reference Z
(reference impedance). Only four items are shown at a time. To view the remaining
items, scroll to the bottom of the screen and continue scrolling.
6. To make a change move the cursor to the desired line and press
ENTER
.
A screen appears in which the present setting can be edited.
7. To store the settings, press
ENTER
.
a) For pH Preamp, choose Analyzer.
b) For pH Reference Z, choose Low.
c) Leave Filter at the default value unless readings are noisy.
For an explanation of terms, see
8. To return to the main display, press
MENU
and then
EXIT
.
5.6
Configuring temperature related settings
5.6.1
Purpose
This section describes how to do the following:
1. Choose temperature units.
2. Choose automatic or manual temperature correction for membrane permeability.
3. Choose automatic or manual temperature compensation for pH.
4. Enter a temperature for manual temperature compensation.
5.6.2
Definitions - chlorine
Automatic
temperature
correction
The free chlorine sensor is a membrane-covered amperometric sensor.
It produces a current directly proportional to the rate of diffusion of free
chlorine through the membrane. The diffusion rate, in turn, depends on
the concentration of free chlorine in the sample and membrane
permeability. Membrane permeability is a function of temperature. As
temperature increases, permeability increases. Thus, an increase in
temperature will cause the sensor current and the transmitter reading
to increase even though the concentration of chlorine remained
constant. In automatic temperature correction, the transmitter uses
Programming the transmitter
Reference Manual
May 2019
00809-0100-3412
56
Emerson.com/Rosemount
Summary of Contents for Rosemount FCL
Page 4: ...4 ...
Page 18: ...Install Reference Manual May 2019 00809 0100 3412 18 Emerson com Rosemount ...
Page 26: ...Wire Reference Manual May 2019 00809 0100 3412 26 Emerson com Rosemount ...
Page 27: ...Reference Manual Wire 00809 0100 3412 May 2019 Rosemount FCL 1056 27 ...
Page 28: ...Wire Reference Manual May 2019 00809 0100 3412 28 Emerson com Rosemount ...
Page 76: ...Digital communications Reference Manual May 2019 00809 0100 3412 76 Emerson com Rosemount ...
Page 107: ...Reference Manual 00809 0100 3412 May 2019 Rosemount FCL 1056 107 ...